Technical Specification
Safety Control  | multi-supervision technology,self-test when turn-on and real-time self-test when infusing  | 
Display  | 320*240 Dot-matrix LCD  | 
Applicable Syringes  | 5ml,10ml,20ml,30ml,50ml,Auto-recognize  | 
Flow Rate Range  | 5ml syringe: 0.1ml/h ~ 200ml/h  | 
10ml syringe: 1ml/h ~ 300ml/h  | |
20ml syringe: 1ml/h ~ 400ml/h  | |
30ml syringe: 1ml/h ~ 600ml/h  | |
50ml syringe: 1ml/h ~1200ml/h  | |
Flow Rate Accuracy  | ±2% (Mechanical Accuracy ±1%)  | 
Limited Volume Range  | 0.1~9999.9ml  | 
Infused Volume Display  | 0.1~9999.9ml  | 
Applicable Syringes Brand  | Preset 15 Kinds,and user added  | 
Purge Rate  | 5ml syringe: 150ml/h  | 
10ml syringe: 220ml/h  | |
20ml syringe: 330ml/h  | |
30ml syringe: 540ml/h  | |
50ml syringe: 800ml/h  | |
Bolus Flow Rate Range  | 5ml syringe: 0.1ml/h ~ 200ml/h  | 
10ml syringe: 1ml/h ~ 300ml/h  | |
20ml syringe: 1ml/h ~ 400ml/h  | |
30ml syringe: 1ml/h ~ 600ml/h  | |
50ml syringe: 1ml/h ~ 1200ml/h  | |
KVO Flow Rate  | 1.0 ~ 5.0ml/h  | 
Body Weight/Drug Mode Rate  | 10 Optional (ml/h,ml/min,mg/h,ug/h,mg/min,ug/min,mg/kg/h,mg/kg/min,ug/kg/h,ug/kg/min)  | 
Occulusion Pressure  | 8 Level Sensitivity,20~140KPa,Dynamic Pressure  | 
Alarms  | Syringe disloacated,Syringe handspike Err.,Pre-alarm of completed limited volume,Completed limited volume,Near empty,Empty,Occulusion,Bolus volume finished,Low battery,Battery empty,Reminder alarm  | 
Power Supply  | AC 100~240V,50/60Hz,Power consumption ≤25VA  | 
Internal Battery  | Ni-MH,12V,800mAh(or Li,11.1V,1800mAh)  | 
Battery Operation  | 3.0(or 7) Hours at 25ml/h  | 
Classication  | Class I, Internal power supply, CF type, IPX4  | 
Data Communication Interface  | wireless (accessories)  | 
Operating Conditions  | Temperature +5℃~+40℃, Relative humidity 20%~95%  | 
Package Size  | 37(W)*24(H)*21(D)cm  | 
Package Weight  | 3kg  |